The Stimulus submenu defines the sweep range in the current channel, depending on the sweep type.
In Cartesian diagrams, the sweep range corresponds to the diagram width and defines the scaling of the x-axis.
In polar diagrams and Smith charts the stimulus axis is lost but the marker functions easily provide the stimulus value of any measurement point.
|
|
|
Start is the lowest value of the sweep variable (e.g. the lowest frequency measured) and corresponds to the left edge of the Cartesian diagram.
Stop is the highest value of the sweep variable (e.g. the highest frequency measured) and corresponds to the right edge of the Cartesian diagram.
Center corresponds to the center of the Cartesian diagram, i.e. (Start + Stop)/2.
Span corresponds to the diagram width, i.e. (Stop – Start).
Power defines the power of the internal signal source.
CW Frequency sets the fixed frequency for the sweep types Power, Time and CW Mode.
START CENTER opens the input field for the Start frequency, if the range is currently defined by Start and Stop. It opens the input field for the Center frequency, if the range is currently defined by Center and Span.
STOP SPAN opens the input field for the Stop frequency, if the range is currently defined by Start and Stop. It opens the input field for the Span, if the range is currently defined by Center and Span.
The Marker
Functions
provide a convenient alternative to the manual entry of the sweep range
parameters. Use the paste
marker list
for convenient entry of Start
and Stop
values.
Some of the Stimulus parameters and the STOP SPAN key are not provided for all sweep types; see Stimulus and Sweep Types.
|
Equivalence |
For a frequency sweep the Start and Stop frequencies or the Center frequency and Span are alternative settings:
|
|
Remote control: |
[SENSe<Ch>:]FREQuency:STARt |
The function and input values of the Start, Stop, Center and Span softkeys depend on the sweep type:
|
Sweep type |
Start [dimension] |
Stop [dimension] |
Center [dimension] |
Span [dimension] |
|
Equivalent hotkey |
|
|
|
|
|
Lin. Frequency |
Start frequency [Hz]
|
Stop frequency [Hz]
|
Center frequency [Hz]
|
Frequency span [Hz]
|
|
Power |
Start power [dBm]
|
Stop power [dBm] |
– |
– |
|
Time |
–
|
Total sweep time [s]
|
–
|
–
|
|
CW Mode |
–
|
Number of points [1]
|
– |
– |
The ranges of numerical values must be compatible with the instrument model.
If a frequency sweep is selected, then the START
CENTER and STOP SPAN are
both toggle keys. In Time or
CW Mode sweep mode the Start, Center and Stop
softkeys are inactive, still you can use the START
CENTER hardkey to select the fixed CW frequency.
|
|
The conditions for the stimulus range depend on the sweep type:
|
The selected sweep range applies to all source ports
and (in the case of a frequency sweep) to all receiver ports of the analyzer.
Use the Port
Configuration
dialog to modify the selected sweep ranges and assign individual ranges
to the analyzer ports.
Opens the numeric entry bar to set the fixed frequency for the sweep types Power, Time and CW Mode.

|
Remote control: |
SOURce<Ch>:FREQuency<Pt>:CW|FIXed |